C H A P T E R
1 6
-
C H A N G E
L O G
V 0.007.246 – 23 APR 2018
Functionality
Implemented filter control in range offsets.
Added a support for a hardware-based watchdog to recover from severe disruption.
Editor results return now waits for the enumeration to finish. This fixes the comms
drop after a sparkover.
Fixed an issue with high frequency class D measurements reporting 0.0000 due to a
timeout.
Tests
IR - Unified the range configuration in DC remainder maintenance to I prove IR test
RLOS - Fixed limits for measure mode for RLOS.
RLOS - Fixed parsing of RLOS measure mode command.
L2 - Trimming is based on the measured ratio rather than the nominal ratio.
L2 - Fixed L2 returning a trimming error on auto secondary signal level.
STRW Fixed bug causing occasional 0.000 results
ILK - Resolved a spurious excessive load error during class D ramp up
TRL - Resolved negative readings for turns ratio by inductance when negative
polarity is measured with no polarity check specified.