10.2.
Accuracy Specifications – Available Tests
10.2.1.
The accuracy is based on the chosen test conditions, and is given by:
where:
where:
where:
The test signal is set according to the value specified as maximum in the test
limits: -
Maximum
Resistance R
Test current
Where:
V = Volts (V)
I = Current (A)
R = Resistance (
AT5600 User Manual 98-119 issue 14

R Test

��
= ��
+ ��
+ ��
+ ��
��
��
��
��
A
is the total accuracy (%)
T
A
is the basic relative accuracy (%)
R
A
is the calibration accuracy (%)
C
A
is the correction for voltage level (%)
V
A
is the correction for current level (%)
I
��
= 0.10%
��
��
= 0.08%
��
0.1%
��
= 0.03% +
��
��
��
0.001%
��
= 0.03% +
��
��
��
V
is the measured voltage (V)
M
I
is the measured current (A)
M
Ω
Ω
<1
1
I=1
I=1/R
Ω)
��
Ω
Ω-
Ω
-10k
10k
50k
V=R*100u
Ω
>50k
V=5