C H A P T E R
1 0
–
S P E C I F I C A T I O N S
10.2.15.
The accuracy of the applied test voltage is 2%.The measurement accuracy is
based on the chosen test conditions, and is given by:
where:
where:
where:
and:
where:
AT5600 User Manual 98-119 issue 14
WATT and STRW Tests
=
+
+
+
A
is the total accuracy (%)
T
A
is the basic relative accuracy (%)
R
A
is the calibration accuracy (%)
C
A
is the correction for voltage level (%)
V
A
is the correction for current level (%)
I
A
is the correction for power factor (%)
PF
= 0.30 % 1.00 %
0.001 %
= 0.80 % + (
0.01 %
= 0.05 % + (
0.01 %
= 0.05 % + (
√ ( 1 −
= 1.0 % ∗ (
F
is the test frequency (Hz)
M
V
is the measured voltage (V)
M
I
is the measured current (A)
M
PF
is the power factor of the load
=
W
is the measured power (W)
VA
is the product of test voltage and magnetizing current
+
)
)
) + (1.0 % ∗ )
2
)
)