15. To select the converter type and a high-side Lo measurement configuration, press:
Notice in this high-side LO, down conversion configuration, the analyzer's source is actuahy
sweeping backwards, as shown in Figure 3-7. The measurement setup diagram is shown in
Figure 3-8.
16. To view the measurement trace, press:
17. To perform a one-sweep power meter calibration over the RF frequency range, press:
Note
Reducing the number of points will turn off the receiver calibration.
The analyzer is now displaying the conversion loss of the mixer calibrated with power
meter accuracy.
Figure 3-7. Diagram of Measurement Frequencies
NETWORK ANALYZER
Figure 3-8. Measurement Setup from Display
reduce the number of points to perform this power meter calibration.