CR 30-X, CR 30-Xm | Introduction to CR 30-X/CR 30-Xm | 37
Cleaning the optical unit
The only maintenance action which you must perform is checking the image
TM
quality. Refer to the User Manual of the NX
Software.
Cleaning the optic unit is required if stripes parallel to the image plate
movement can be seen in the image. When you recognize this type of artefact,
when using the digitizer, clean the optic unit using the cleaning brush.
To clean the optic unit, proceed as follows:
1. Open the cassette unit.
2. Take out the cleaning brush.
3. Open the lid positioned at the right side.