Reading the Initialization Data in a Configuration with Fast ID

1. Check that the system is ready for operation:
The status indicator on the digitizer constantly lights up in green.
2. Click on Read and Initialize Cassette (1) in the Functionality Overview
pane of the Main Menu window of the NX station.
The Read and Initialize Cassette pane (2) is opened in the middle section
of the Main Menu window:
1
For more information, refer to the NX Key User Manual, document 4421.
3. Click on the Read button at the NX workstation.
The digitizer waits for the cassette and the status indicator constantly
lights up in green.
4. Insert the cassette containing the image plate into the cassette slot [1] of
the digitizer as shown below.
Make sure to insert the cassette with the black side to the top and with the
shutter opening mechanism and the locking mechanism inside the
digitizer.
Make sure that the cassette is firmly pushed up to the right side of the slot
[2]. Otherwise, the digitizer can not read the image plate.
CR 30-X, CR 30-Xm | Operating CR 30-X/CR 30-Xm | 75
2
2386I EN 20151020 1518