CR 30-X P
C
LATES AND
ASSETTES
Cleaning
Cleaning the Tray and Image Plate
When to clean the tray and image plate?
The inner lining of the CR 30-X cassette is made of fleece. This ensures a high
degree of protection against electrostatic charging and dust collection on the
image plates. Nonetheless it is recommended to clean the tray/image plate
•
At least once every month or whenever there is doubt about particles
visible on the X ray images.
•
More frequent cleaning is also required in case of excessive dust or under
very dry conditions. (The screen cleaner contains antistatic agents which
reduce buildup of static charges and dust collection.)
Permitted cleaning products?
•
Use only AGFA CR Phosphor Plate Cleaner and a soft lint-free cellulose
cloth to clean the image plate.
•
Use only a dry cloth or blow away possible dust with compressed air to
clean the tray.
18
Note: Please make sure that the tray and image plate set always goes into the
same cassette when cleaning. It is important that an image plate is not put
together with a tray initialized for another image plate.
2387A EN 20060913