3. TABLE MAINTENANCE

3.1. CLEANING AND DISINFECTING

Warning: always disconnect the equipment from the mains before cleaning or disinfecting it.
Do not allow water or other liquids to enter the equipment as such liquids may cause
short-circuits or corrosion.
Clean all enamelled parts with mild detergent only. Apply with a clean cloth, and then rub down with a
dry woollen cloth.
Never use abrasive, solvent or corrosive cleaning agents.
Clean chrome parts by rubbing with a dry woolen cloth. Should you desire to polish such parts, use car wax.
Do not use abrasive polishes.
Wiping them with a cloth moistened with a disinfectant solution may disinfect the equipment,
accessories and connecting cables.
Never use solvent, corrosive or phenol-based disinfectants.
The table and control console must not be exposed to gaseous disinfectants.
Spray disinfectants are not recommended since they can enter the equipment causing short-circuit
and/or corrosion.
If the procedure room containing the equipment is to be disinfected by mean of an atomiser, it is
mandatory to respect the following recommendations:
After the mist has dispersed, the covers may be removed and the equipment disinfected by wiping in
the manner described above.
The equipment must not be used in the presence of disinfectants that vaporize, forming potentially
explosive gases. All vapour must be allowed to disperse before returning the equipment to use.
Disinfecting procedures must comply with all current and relevant regulations and recommendations,
including those concerning prevention of potentially explosive hazards.
Digital Remote Controlled Table with dynamic flat panel detector
 Switch off the equipment
 Allow them to cool well in advance
 Cover them carefully with plastic or similar sheeting.