Equipment Classification

This device is classified as following:
Table 1: Equipment classification
Class I equipment
Type B equipment
Water ingress
Cleaning
Disinfection
Flammable anesthetics
Operation
CR 30-X, CR 30-Xm | Introduction to CR 30-X/CR 30-Xm | 15
Equipment in which protection against electric
shock does not relay on basic insulation only, but
includes a power supply cord with protective earth
conductor. For earth reliability always plug the
main power cord into an earthed mains power
outlet.
Not classified.
The patient does not get in contact with any part of
the equipment.
This device does not have protection against
ingress of water.
See section on cleaning and disinfecting.
See section on cleaning and disinfecting.
This device is not suitable for use in the presence of
a flammable anesthetic mixture with air, or in
presence of a flammable anesthetic mixture with
oxygen or nitrous oxide.
Continuous operation.