44 | CR 30-X, CR 30-Xm | Introduction to CR 30-X/CR 30-Xm

Quality Control

If no specific regulations are valid, a regular quality control by means of the
Agfa Auto QC2 tool is required at least once a month to maintain a safe and
effective system.
For mammography, Agfa recommends to use the "Routine Quality Control
Tests for Full Field Digital Mammography Systems" document, created by the
NHSBSP (National Health Service Breast Screening Program, UK).
2386I EN 20151020 1518
WARNING:
Unnoticed image quality degradation can cause false
negative diagnosis.
Apply regular quality control according local regulations.