AGFA CR 30-X User Manual page 81

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

Condition
Connection
problem
between
digitizer and
the Digitizer
Remote
Display.
Connection
problem
between
digitizer and
NX PC.
Errors During Operation
If errors occur during operation, you can consult the Digitizer Remote Display
messages on the control PC. The Digitizer Remote Display is independent
from the NX Software.
Condition
Empty
cassette (no
image plate in
cassette)
Empty
cassette (no
image plate in
cassette) or
image plate
jam.
CR 30-X, CR 30-Xm | Operating CR 30-X/CR 30-Xm | 81
Message at
Status
Digitizer
indicator
Remote
Display
No error
Red blinking
message on NX
fast
PC.
Red blinking -
3 pulses
Control PC message
Cassette error
"Empty cassette
found at start-up.
Remove cassette."
"Open digitizer to
check for image plate
jam or check for an
empty cassette and
restart the digitizer."
Action
Check if Digitizer
Remote Display is
running. Start/restart
Digitizer Remote
Display.
Check the Ethernet
cables. If the error
remains, restart PC and
digitizer or call service.
Status
indicator
Red,
Remove the cassette
flashing
Switch off the digitizer
and open the front
cover. If there is no
image plate visible in
the transport unit
close and switch on
the digitizer. After
Red,
restart remove the
constant
cassette from the
digitizer and check if
there is an image plate
inside the cassette.
In case of an image
plate jam remove the
2386I EN 20151020 1518
Action