• If you don't operate the digitizer correctly or if you don't have it serviced
correctly, Agfa is not liable for resulting disturbances, damages or injuries.
• If you notice conspicuous noise or smoke, disconnect the digitizer
immediately.
• Do not pour water or any other liquid over the device.
• The digitizer complies with the EN 60601-1 and UL 60601-1 standards for
Medical Electrical Equipment. This means that, although it is absolutely
safe, patients may not come in direct contact with the equipment.
Therefore the operator console must be placed outside a radius of 1.5 m
(EN) or 1.83 m (UL) around the patient (according to the local valid
regulation).
1
Patient environment: R = 1.5 m (1.83 m)
1.
Patient environment: h = 2.5 m (2.29 m)
2.
• Position the digitizer in a way, that it could be easily unplugged in order to
achieve separation from mains.
• Perform no other operations on the digitizer than those described in this
document.
• Switch the system off before moving it. When reaching the new position,
switch the system on again.