AGFA CR 30-X User Manual page 83

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

Condition
Control PC message
present at the PC and
re-scan the cassette, if
necessary. If the
image is present, the
cassette was not
erased possibly. Erase
the cassette again in
any case."
Digitizer
cannot detect
the cassette
chip on the
"Cassette chip not
image plate
readable or cassette
tray e.g. tray
tray is missing.
is missing or
Remove cassette."
was not
inserted
correctly.
"Image transmission
Communicati
failed. System is
on error,
retrying. If image
Ethernet cable
does not appear,
is not plugged
please restart the
in
PC."
"ID button not
Time-out after
pressed within the
5 minutes if
timeout period.
the ID button
Image plate was not
of the image
scanned. Cassette
processing
stays clamped until
software was
ID data are supplied
not pressed.
by the operator."
Messages will stay active until the problem is solved or the dialog box message
on the control PC is confirmed by clicking the confirm button.
Note: For error messages not listed in the table above, please refer
to the instructions within the displayed error message text.
CR 30-X, CR 30-Xm | Operating CR 30-X/CR 30-Xm | 83
Status
indicator
Red,
flashing
After
unlocking:
green,
constant
Red,
flashing
After ID
button has
been
pressed:
green,
flashing
Action
In order to release the
cassette, confirm the
message.
Open the cassette and
check if the tray is
inserted correctly or
use another cassette
and call service.
Check if Ethernet
cable is plugged in at
PC and digitizer.
Visually check if the
Ethernet cable is
damaged. If the error
remains restart PC and
digitizer or call
service.
Confirm the message
on the PC and press
the ID button of the
image processing
software.
2386I EN 20151020 1518