50 | CR 30-X, CR 30-Xm | Getting started
plate of the cassette inserted next. If no cassette with image plate has been
inserted after 60 seconds, the system automatically returns to standby mode.
Status Indicator
The indicator informs the user by light signals about the status of the digitizer.
It is positioned at the front of the digitizer, so that it is visible from a distance.
The indicator is divided in two parts. The upper part is used to inform the
operator about the image plate erasing cycle progress and is only then lit. The
lower part is used for all other operational indications.
Blue
1.
Green or red
2.
Color
Blue
Green
Red
Related Links
Troubleshooting Checklist
2386I EN 20151020 1518
1
2
Constant/
Blinking
Constant
Constant
Flashing
Constant
Blinking
Blinking fast
Blinking - 3
pulses
Status
Activating the erasing
cycle
• Standby mode
(Ready)
• Cassette is ready for
removal
Busy with scanning,
erasing and return of
the IP into cassette
Service mode
• Warm up / Self-test
• Processing Software
down
• Error
Digitizer not connected
to Digitizer Remote
Display UI
Digitizer not connected
to control PC
on page 80
Action
• Proceed.
• Remove cassette.
Wait.
Check workstation for
further information and
detailed instructions.
Refer to section
'Troubleshooting'.