Top Tray; Universal Microscope Carrier - Leica HistoCore AUTOCUT Instructions For Use Manual

Rotary microtome
Table of Contents

Advertisement

Caution
When using super mega-cassette clamp, the orientation is not set to position "0" when directional
fixture for specimen clamps is used with the rigid knife holder base, or the back-lighting illumination
is attached.
The instrument may malfunction leading to delayed diagnosis.
• The orientation must be in position "0" and the cover for backlighting illumination must be detached.
• NEVER use the super mega-cassette clamp with backlighting illumination.
• The force balance system has to be adjusted when using the super mega cassette clamp.
6.6

Top tray

Fig. 91
6.7

Universal microscope carrier

Tips
Unpack all accessories in the package and check for completeness.
HistoCore AUTOCUT
The top tray is mounted on the hood of the
microtome. The mark
the top tray and the hood helps to orient the
installation of the top tray.
The top tray is for storage of the utensils used
during sectioning as well as the sectioned
specimens.
1
Optional Accessories
(→ "Fig. 91-1")
on both
6
83