Leica HistoCore AUTOCUT Instructions For Use Manual page 111

Rotary microtome
Table of Contents

Advertisement

Two-in-one blade holder E
1
2
Fig. 141
1. Remove the insert
2. Remove the pressure plate first
blade clamping lever
bolt
(→ "Fig. 141-7")
3. Remove the clamping top. To do so, rotate the clamping lever
displacement counterclockwise and pull it out sideways; push the clamping top
until it can be removed from the segment arc
4. Loosen the eccentric bolt using Allen key No. 4 and remove the segment arc from the blade
holder base.
5. Clean all parts of the two-in-one blade holder E.
Caution
During cleaning, the parts of blade holders are mixed up.
Poor section quality.
• Do not mix up blade holders during cleaning.
HistoCore AUTOCUT
9
3
7
6
4
5
7
6
9
(→ "Fig. 141-9")
for low-profile blades.
(→ "Fig. 141-4")
(→ "Fig. 141-6")
counterclockwise and pull it out sideways; pull out the
and take the pressure plate off.
Cleaning and Maintenance