6.1.2 Directional fixture for specimen clamps

1
2
3
4
Fig. 69
7

6.1.3 Fine-directional fixture for specimen clamps

1
3
Fig. 70
HistoCore AUTOCUT
5
8
4
2
• Loosen the eccentric bolt
turning it counterclockwise.
• Completely unscrew the thrust piece
69-2")
with a flat-tip screwdriver and pull it out
with spring
(→ "Fig. 69-3")
• Completely unscrew the setscrews
and
(→ "Fig.
6
• Install the directional fixture for specimen clamps
by tightening the screws in the bores
69-8")
(2 screws are accessible through the bores)
and screw them in using an Allen key No. 3. Insert
the screws
(→ "Fig. 69-7")
and screw them in using an Allen key No. 3.
• Insert the spring
69-4")
with the flatter side into the thrust piece
(→ "Fig.
69-2"). Completely screw in the thrust
piece with a flat-tip screwdriver.
• Completely screw in the setscrews
(→ "Fig.
69-6").
• Before the fine-directional fixture for specimen
clamps can be mounted, loosen 4 screws
(→ "Fig. 70-2")
remove the fixture for specimen clamps from
the baseplate
• Using the 4 supplied screws
the Allen key No. 3, fasten the baseplate to the
object head
• Now, screw the fine-directional fixture for
specimen clamps with the 4 screws
70-2")and the Allen key No. 3 onto the object
head.
Optional Accessories
(→ "Fig. 69-1")
and pin
(→ "Fig.
(→ "Fig. 69-5")
69-6").
in the bores as shown
(→ "Fig. 69-3")
and pin
(→ "Fig. 69-5")
(Allen key No. 3) and carefully
(→ "Fig.
70-1").
(→ "Fig. 70-3")
(→ "Fig.
70-4").
6
by
(→ "Fig.
69-4").
(→ "Fig.
(→ "Fig.
and
(→ "Fig.
69