Inserting The Two-In-One Blade Holder E; Adjusting The Clearance Angle - Leica HistoCore AUTOCUT Instructions For Use Manual

Rotary microtome
Table of Contents

Advertisement

3. The blade holder base
plate. This allows bringing the two-in-one blade holder E to optimal sectioning position in
relation to the specimen. To secure the blade holder base, turn the clamping lever
53-1")
clockwise.

5.2.2 Inserting the two-in-one blade holder E

Fig. 54
1. Loosen the eccentric bolt
2. Place the two-in-one blade holder E
(→ "Fig. 54-7")
3. To clamp, retighten the eccentric bolt
5.3

Adjusting the clearance angle

2
1
Fig. 55
The index marks (0°, 5° and 10°) for adjustment of the clearance angle
right side of the two-in-one blade holder E
the right side of the blade holder base
the clearance angle.
1. Loosen the screw
blade holder E
HistoCore AUTOCUT
(→ "Fig. 53-2")
can be moved back and forth on the microtome base
(→ "Fig. 54-9")
(→ "Fig. 54-8")
of the knife holder base
3
(→ "Fig.
(→ "Fig. 55-1")
(→ "Fig. 55-3")
using an Allen key No. 4
(→ "Fig. 55-2")
can be moved.
8
using an Allen key No. 4
with the underside groove onto the T-piece
(→ "Fig.
53-2").
(→ "Fig.
54-9").
55-2"). There is also an index mark
which serves as a reference point when adjusting
(→ "Fig. 55-6")
Operation
(→ "Fig.
7
10
9
(→ "Fig.
54-10").
4
5
6
(→ "Fig. 55-4")
are located on the
(→ "Fig. 55-5")
until the two-in-one
5
on
57