Leica HistoCore AUTOCUT Instructions For Use Manual page 74

Rotary microtome
Table of Contents

Advertisement

6
Optional Accessories
Fig. 76
Warning
The remaining broken edge of cassette lid may lead to poor section quality due to insecure clamping.
Serious injury of persons.
• When using a cassette whose lid is molded on, make sure that the broken edge left by removing the
lid does not prevent the specimen from being securely clamped – if necessary, the specimen must be
clamped horizontally.
Caution
When using thin-walled cassettes, the cassettes become deformed or insecurely clamped into place,
or other problems because of the clamping system.
Damage to specimen/Delayed diagnosis.
• Be careful when using thin-walled-cassettes. Make sure that the thin-walled-cassettes are clamped
securely in place.
• If the user tries clamping the cassette and realizes that it is not securely clamped into place, use a
more stable one.
Caution
Paraffin deposits on the outside of the cassette can make the universal cassette clamp dirty.
The dirt prevents the cassette from clamping in securely and can lead to sections being too thick or
thin, chattering within the section and, in the worst-case scenario, damage to the specimen.
• Prior to sectioning, the user has to verify that the specimen is clamped securely.
• Remove paraffin deposits from the universal cassette clamp.
74
1
2
Surgipath cassettes with minimum dimensions of
39.8 x 28 mm and maximum dimensions of 40.9 x
28 mm may be clamped in the universal cassette
clamp (UCC) horizontally as well as vertically.
• Pull the lever
(→ "Fig. 76-1")
operator.
• Mount the cassette
(→ "Fig. 76-2")
or vertically as required.
• To clamp the cassette, release the lever
(→ "Fig.
76-1").
towards the
horizontally
Version 1.1, Revision B