Leica HistoCore AUTOCUT Instructions For Use Manual page 33

Rotary microtome
Table of Contents

Advertisement

Warning
Instrument not positioned safely with all 4 feet on an appropriate laboratory table.
Serious injury of persons and/or property damage.
• Put the instrument only on a stable, vibration-free laboratory table with horizontal, flat table top.
Ground needs to be vibration-free as far as possible.
• Always make sure that all 4 feet are positioned completely on the laboratory table.
• If the instrument has been potentially moved by other persons (for example for servicing) always
double-check correct positioning.
Caution
Accessories/components loosening/damaged during transportation.
Property damage or delayed diagnosis.
• The packaging has two indicators, ShockDot Impact Indicator and Tilt Indicator, that indicate
improper transport. When the instrument is delivered, check these first. If one of the indicators
is triggered, the package was not handled as prescribed. In this case, please mark the shipping
documents accordingly and check the shipment for damage.
Tip
The transport carton and included retaining elements should be kept in case a return shipment is
necessary later. To return the instrument, follow the instructions below in reverse order.
Fig. 11
HistoCore AUTOCUT
Setting up the instrument
1
• Remove the packing strap and the adhesive
2
tape
(→ "Fig.
• Remove the carton lid
11-1").
(→ "Fig.
11-2").
4
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1405195647214 9bio000c214 9auto00c1

Table of Contents