Markings On The Instrument Itself; Transport And Installation - Leica HistoCore AUTOCUT Instructions For Use Manual

Rotary microtome
Table of Contents

Advertisement

2.2.1 Markings on the instrument itself

Warning
Failure to follow the correct operating instructions (as defined in the Instructions for Use) of the
corresponding markings and warning triangles on the instrument.
Serious injury of persons and/or damage to instrument, accessories or specimen.
• Pay attention to the markings on the instrument and strictly follow operating instructions
described in the Instructions for Use when operating or replacing the item marked.

2.2.2 Transport and installation

Warning
Accessories/instrument falling out of the package during unpacking.
Serious injury of persons and/or property damage.
• Be careful when unpacking the instrument.
• Once unpacked, the instrument must be moved only in an upright position.
• Thoroughly follow the instructions given in the Unpacking Instructions attached at the outside of
the package or the actions described in the Instructions for Use.
Warning
Instrument is lifted inappropriately.
Serious injury of persons and/or property damage.
• When lifting the instrument touch it only at the lifting points described in the Instructions for Use
and Unpacking Instructions (base plate at the front and under the instrument on the back).
• Never lift the instrument by the handwheel grip, electric coarse feed wheel, or object head.
• Always remove the section waste tray before transporting the instrument.
Warning
Careless moving of the instrument.
Serious injury of hands and/or fingers due to getting crushed between instrument and working
surface.
• When moving the instrument, touch it only at the lifting points described in the Instructions for Use
and Unpacking Instructions (base plate at the front and under the instrument on the back).
• Thoroughly pay attention that hands do not get between instrument and working surface.
Warning
Instrument not positioned safely with all 4 feet on an appropriate laboratory table.
Serious injury of persons and/or property damage.
• Put the instrument only on a stable, vibration free laboratory table with horizontal, flat table top.
Ground needs to be vibration-free as far as possible.
• Always make sure that all 4 feet are positioned completely on the laboratory table.
• If the instrument has been potentially moved by other persons (for example for servicing) always
double-check correct positioning.
HistoCore AUTOCUT
2
Safety
11