Leica HistoCore AUTOCUT Instructions For Use Manual page 85

Rotary microtome
Table of Contents

Advertisement

12
13
Fig. 95
2
Fig. 96
Warning
The universal microscope carrier tips.
Serious injury of persons.
• After installing the vertical column, immediately place the microtome onto the baseplate so that the
front feet of the microtome are positioned in the shallow recesses
Tips
The lock washer must be positioned between the baseplate and vertical column to prevent
unintentional turning of the column.
HistoCore AUTOCUT
• Attach vertical column. Insert the cap screw
(→ "Fig. 95-12")
from below. Place the lock washer
95-13")
the silver vertical column
the baseplate from above and tighten with the
Allen key size 8.
19
Optional Accessories
into the hole of the baseplate
on the cap screw from above. Thread
(→ "Fig. 96-2")
(→ "Fig.
96-19").
6
(→ "Fig.
onto
85