Four-digit display on the instrument control panel
3
Fig. 26
Tip
When the instrument is switched off using the main power switch, both values (section thickness
sum and section number) are erased from memory.
Specimen retraction
To prevent damage to the knife and specimen, the specimen is moved away from the knife during the
return motion to the upper home position of the object head.
In motorized sectioning mode, the retraction depends on the setting of the sectioning speed.
In manual mode, the retraction value can be selected in 5 µm increments between 5 and 100 µm.
Specimen retraction is set to 10 µm at the factory.
The specimen retraction can also be deactivated for the manual and motorized operation if required.
The selected setting is maintained when the instrument is turned off.
Configuring the retraction settings
1
3
4
6
7
Fig. 27
HistoCore AUTOCUT
The value showing on the four-digit display is adjustable.
When the
Σ-µm
1
section thicknesses in µm for all sections completed since the instrument
was switched on.
When the
LED
Σ-n
previously completed sections.
2
• To change the display mode, push
the LED of the desired mode is illuminated.
4
• Press
button
CLEAR
number.
• This will only reset the currently displayed value.
• To call up the retraction settings, press the
CLEAR
(→ "Fig. 27-2")
2
• The current set value is displayed as a three-digit number in the four-digit
display, e.g. 025 = 25 µm
• Select the desired retraction value. The retraction value can be adjusted in
increments of 5 µm to a maximum of 100 µm using buttons
(→ "Fig. 27-5")
on the separate control panel. The setting can be off too.
5
• To exit the retraction settings, press
retraction movement takes place by the newly selected value after each
section.
• While the specimen is in retraction, the yellow LED
RETRACT
display lights up.
LED
(→ "Fig. 26-1")
is lit, the display shows the sum of the
(→ "Fig. 26-2")
is lit, the display shows the number of all
MENUMODE
(→ "Fig. 26-4")
to reset section thickness sum or section
buttons simultaneously.
(→ "Fig.
27-3").
MENUMODE
Operation
button
(→ "Fig. 26-3")
(→ "Fig. 27-1")
MENUMODE
(→ "Fig. 27-4")
(→ "Fig.
27-6"). A
(→ "Fig. 27-7")
on the
5
until
and
or
47

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1405195647214 9bio000c214 9auto00c1

Table of Contents