Leica HistoCore AUTOCUT Instructions For Use Manual page 61

Rotary microtome
Table of Contents

Advertisement

Removing the blades
1
3
Fig. 61
1. Rotate the clamping lever
2. Push the pin
Tips
Use the blade ejector to safely eject the blade.
3. Bring the safety guard
remove the blade from the right side and out.
Once the blade has been removed from the two-in-one blade holder E, it is disposed in the used blade
container on the downside of the blade dispenser
Fig. 63
Warning
Knives or blades are handled and/or disposed in an inappropriate manner.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Work especially thoroughly and carefully when handling knives and/or blades.
• Always wear dedicated clothes (including cut-resistant gloves) when handling knives and/or
blades.
• Dispose knives and blades always in a safe place (for example, dedicated knife case) and an
appropriate manner that ensures that people cannot get hurt.
• Never place a knife anywhere with the cutting edge facing upwards and never try to catch a
falling knife.
• Always cover the edge of the knife/blade with the safety guard before clamping the specimen.
HistoCore AUTOCUT
2
(→ "Fig. 61-2")
(→ "Fig. 61-3")
on the blade ejector
(→ "Fig. 62-1")
1
Fig. 62
counterclockwise.
.
down. Use the brush with magnet
(→ "Fig.
63").
5
Operation
4
2
(→ "Fig. 62-4")
to
61