Leica HistoCore AUTOCUT Instructions For Use Manual page 45

Rotary microtome
Table of Contents

Advertisement

Trimming mode
3
1
2
4
Fig. 24
One memory position
Fig. 25
Tip
While the object head is moving to the memory position, the movement can be stopped by pressing
the
MEMO
button or any other buttons, or by rotating the electronic coarse feed wheel.
The
button is enabled between the front limit position (not included) and home position (not
MEMO
included).
HistoCore AUTOCUT
In the trimming mode, the coarse feed buttons operate a continuous
movement as long as the button is held down. The double-arrow button for
fast coarse feed backward movements has a lock-in function.
• To start the fast backwards movement (away from the knife) press the
button
(→ "Fig.
moved to the rear end position.
• To stop the movement, press any of the four coarse feed buttons.
• The yellow LED
head is in motion, and remains lit continuously when the rear end position is
reached.
• Press the button
travel continues as long as the button is held pressed.
• Press the appropriate button to start a rapid or slow forward movement.
The travel continues as long as the button is held pressed.
• During the forwards movement, the yellow LED
flashes. When the front end position is reached, an acoustic signal is heard
and the LED stops flashing and remains lit.
Set one memory position using the
1
separate control panel.
To set one memory position, move the object head to the desired position,
and press the
MEMO
2
indicates
(→ "Fig.
SEt
up in green.
3
If one memory position is stored, after pressing the
head moves to the memory position; a short beep is heard and the three-digit
display indicates
position.
4
To change the stored memory position, move the object head to the new
position, and press on the
5
is heard; the three-digit display indicates SEt; the green LED of the
button keeps lighting on.
To clear the memory position, press on the
second. Two short beeps are heard; the three-digit display indicates
(→ "Fig.
25-5"); the green LED of the
24-1"). After the button is pressed, the specimen head is
(→ "Fig. 24-2")
in the button flashes while the specimen
(→ "Fig. 24-3")
to start the slow backwards movement. The
MEMO
button. A short beep is heard; the three-digit display
25-2"); the LED
(→ "Fig. 25-3")
At
(→ "Fig. 25-4")
when the object head reaches the memory
MEMO
button for about 1 second. A short beep
MEMO
Operation
(→ "Fig. 24-4")
in the button
button
(→ "Fig. 25-1")
on the
of the
button lights
MEMO
MEMO
button, the object
button for more than 3
MEMO
button is greyed out.
5
MEMO
CLr
45