HistoCore AUTOCUT
Turn the instrument on with the power switch at the rear.
I = ON; O = OFF
This is followed by a beep.
The instrument initializes.
The display (here only as an example) of the software version
follows in the four-digit LED display. This display disappears after 2
seconds and "00.00" is displayed. After the microtome is switched
on, the display fields and LEDs of all activated functions are lit up on
the instrument's control panel and display field.
The three-digit LED display indicates the last value set for the
section thickness or trimming thickness, depending on which
setting was last activated. This is indicated simultaneously on the
control panel and microtome. The LED of the active mode (here the
section thickness) is illuminated in green.
When the red LED in the
instrument is illuminated, either
• the emergency-stop function has been activated (the emergency-
stop button is pressed or the foot switch depressed); or
• the dummy plug (or the optional foot switch) is not correctly
connected to the jack, or not connected at all.
If the yellow LED in the
mechanical handwheel lock or the handwheel brake is activated.
If the yellow LED in the
the electronic brake is activated.
The instrument cannot be used as long as the LED is lit or flashing.
Setting up the instrument
field on the control panel of the
E-STOP
M-STOP
field of the control panel is lit, the
field of the control panel is flashing,
M-STOP
4
39