Warning
Emergency stop does not work after software crash.
Specimen damage.
• Try to restart the instrument.
• If restart fails, unplug the power cord and contact your Leica service representative.

5.1.4 Electronic coarse feed wheel

Fig. 48
1
3
Fig. 49

5.1.5 Fine directional specimen holder fixture

Tip
In the quick clamping system of the directional specimen holder fixture, all specimen clamps
available as optional accessories can be used.
HistoCore AUTOCUT
The coarse motion serves for a fast horizontal
forwards movement of the specimen - towards
the knife - and backwards - away from the knife.
The rotation of the electronic coarse feed
wheel
1
counterclockwise to advance the object head.
• To set clockwise rotation to advance the object
head, switch the instrument on while holding
the slow-forward coarse feed button
49-1")
display indicates
2
• To set counterclockwise rotation to advance
the object head, switch on the instrument and
press the slow-backward coarse feed button
4
(→ "Fig.
for counterclockwise
The settings of the electronic coarse feed wheel
rotation is shown for about 4 seconds on the
three-digit display when the instrument switches
on. The factory-set rotation direction of the
electronic coarse feed wheel is clockwise.
(→ "Fig. 48-1")
can be set as clockwise or
on the separate control panel; three-digit
for clockwise
C
49-3"); three-digit display indicates
(→ "Fig.
5
Operation
(→ "Fig.
(→ "Fig.
49-2").
CC
49-4").
53