Operation

15 minutes alarm

To remind the user to monitor the core temperature con-
stantly at higher heater output levels,
– an audible alarm is emitted every 15 minutes at heater
level 4 and above. At the same time:
1 the yellow LED flashes.
– The red central alarm lamp on the radiant warmer starts to
flash.
2 The heater level display flashes and
3 the following message appears in the text display:
»15 min. patient temp check required
Press OK button to acknowledge alarm«.
G
4 Press
button
or
5 Press OK button.
The audible alarm will stop and the yellow LED and the
central alarm lamp will go out. The heating will remain
on. The heating level display will be continuously lit.
If there is no acknowledgement after 15 minutes the heat-
ing will switch off.
– The audible alarm remains on.
1 Yellow LED flashes and
– The red central alarm lamp on the radiant warmer flashes.
2 The display of the heating level selected will disappear and
3 the following message will appear in the text display:
»15 min. patient temp check required
Press OK button to acknowledge alarm«.
G
4 Press
button
or
5 Press OK button.
The heating will be switched on again.
The audible alarm will stop and the yellow LED and the
central alarm lamp will go out. The heating level display
will be lit again.
If a skin temperature sensor is connected or attached in
manual mode, it will generate a skin temperature display but
will not affect the heater output.
24
3
10
5
2
1
skin C
manual
mattress . C
skin
man .
Babytherm 8010
1
4
15 min
C
< 36 C
5