3. Basic Safety Measures

General Safety Measures
Ground the instrument before operation!
Do not disassemble any part of the instrument. Disassembling of the product is
permitted only to persons certified by NT-MDT Co.
Do not connect additional devices to the instrument without prior advice from
NT-MDT Co.
The instrument contains precision electro-mechanical parts. Therefore, protect it from
mechanical shocks.
Protect the instrument against the influence of extreme temperature and moisture.
For transport, provide proper packaging of the instrument to avoid its damage.
Electronics
To reduce possible influence of power line disturbances on the measurements, use of a
surge filter is recommended for electrical supply of the instrument units.
Before operation, set the voltage selector of the SPM controller to the position
corresponding to local electrical power line (this is only done with the controller
being off!).
Switch the instrument off before any operation on connecting/disconnecting its cable
connectors. Disconnecting or connecting the cable connectors during operations may
cause damage to the electronic circuit and disable the instrument. A warning label is
attached to the SPM controller of the instrument (Fig. 3-1).
Laser
Observe safety measures for operation with devices containing sources of laser
radiation. See warning labels on the laser sources and on AFM measuring heads
(Fig. 3-2).
20
Fig. 3-1