NTEGRA Spectra Probe NanoLaboratory. (Inverted Configuration with Renishaw Spectrometer).
3. Withdraw the probe from the sample through the following sequence of operations:
a. Go to the
Approach
(Fig. 9-36).
b. Double-click in the
Control panel of the
of 2÷3 mm.
c. To retract the probe from the sample, click the
group (Fig. 9-38).
4. Retract the lens from the sample by rotating the manual approach knob clockwise.
5. Switch off the laser.
6. Switch off the SPM controllers.
7. Switch off the spectrometer.
8. Close the Control program.
50
tab with the
Fig. 9-36. Control panel of the
input field of the
Moving
tab to open a slider. Position the slider at the value
Approach
Fig. 9-37
Fig. 9-38
button in the Main Operations panel
Approach
tab
group (Fig. 9-37) on the
Backward
button of the
Backward