NT-MDT NTEGRA Spectra Instruction Manual page 35

Inverted configuration with renishaw spectrometer
Hide thumbs Also See for NTEGRA Spectra:
Table of Contents

Advertisement

4. Define cooling temperature of the CCD-matrix (in the range of –40÷–50 °C) in the
input field available in the Toolbar (Fig. 9-5).
Set T(°C)
5. Start cooling the CCD-matrix by pressing the
of the input field. The cooling process will begin and the text caption of the pressed-in
button will indicate the current temperature of the matrix.
6. Adjust parameters controlling spectrum detection by the CCD-camera through the
following steps:
a. Enter the exposure time in the range of 0.2÷1 s in the
b. Define the highest rate for spectra detection by selecting
group of the Instruments parameters adjustment panel (Fig. 9-4).
Shift Speed
c. To acquire a spectrum from the whole CCD-matrix, click the
d. Collect the spectrum (Fig. 9-6) by clicking the
Fig. 9-6. Spectrum collected from the whole CCD-matrix
NOTE. To activate the option of scan-time updating the spectrum image,
click the
NOTE. A spectrum to be saved should be appended to the current data file as a
separate frame using the
To improve the signal/noise ratio at low levels of the signal, the spectrum is collected only
from a limited region of the CCD-matrix that collects the major portion of the radiation:
1. To receive the image from the CCD-matrix, press the
2. Acquire the scan by clicking the
the signal is concentrated (Fig. 9-7).
Chapter 9. Measuring with the Confocal Microscopy Mode
Fig. 9-5. Toolbar
button.
Toolbar button.
button. The collected image displays where
button placed to the right
input field.
Exposure time
in the
list on the
1
H
button.
button.
button.
35