NT-MDT NTEGRA Spectra Instruction Manual page 21

Inverted configuration with renishaw spectrometer
Hide thumbs Also See for NTEGRA Spectra:
Table of Contents

Advertisement

Scanner
Do not press the scanner more than it is needed to install the probe. Avoid scanner
shocks and its lateral deformation. Remember that the width of scanner walls is
only 0.5 mm.
Optical components
Optical components include the following units and elements:
optical elements in the spectrometer unit;
laser.
To avoid contamination of the optical elements with dust or moisture, keep the room clean
and never disperse sprays or other substances near the instrument. Do not touch the optical
elements. For cleaning the optical components, use special cleaning sets.
ATTENTION! Never uncover the spectrometer without extreme necessity. Only
an authorized person is permitted to clean elements of the spectrometer. In the
case of contamination, request for help to the manufacturer.
Radiation detectors
When storing or handling the CCD-camera, avoid ingress of contamination into the
camera. Before connecting the water cooling system, turn off the computer and disconnect
the camera from the computer. Prevent ingress of moisture into the camera case. Keep the
vent holes on the case open when using the camera. Prevent too intensive illumination of
the camera. Make sure that the camera is at the room temperature when finishing the work.
Chapter 3. Basic Safety Measures
Fig. 3-2
21