NT-MDT NTEGRA Spectra Instruction Manual page 43

Inverted configuration with renishaw spectrometer
Hide thumbs Also See for NTEGRA Spectra:
Table of Contents

Advertisement

Approaching Probe to Lens Optical Axis
1. Approach the sample to the probe by contact or semicontact mode. Details are
explained in Performing Measurements, part 3.
2. Retract the probe from the sample to distance of 0.1÷0.2 mm through the following
steps:
a. Double-click in the
Control panel of the
0.1÷0.2 mm.
b. To retract the probe from the sample, click the
3. Move the probe to the laser spot on the sample (Fig. 9-24) with the moving screws of
the measuring head (Fig. 9-25).
1 – screw for moving along X direction; 2 – screw for moving along Y direction
Chapter 9. Measuring with the Confocal Microscopy Mode
input field of the
Moving
tab to open a slider. Position slider at the value of
Approach
Fig. 9-22
Fig. 9-23
Fig. 9-24. The cantilever approached to the laser spot
Fig. 9-25. Measuring head moving screws
group (Fig. 9-22) on the
Backward
button (Fig. 9-23).
43