NT-MDT NTEGRA Spectra Instruction Manual page 17

Inverted configuration with renishaw spectrometer
Hide thumbs Also See for NTEGRA Spectra:
Table of Contents

Advertisement

Chapter 2. Design
This connection needs additional cables, a switching cable and a Z-splitter, which
distribute signals from the scanners of the measuring head, of the exchangeable mount, and
of the lens.
The Z-splitter separates signals from the slave controller to different scanners. Z-socket of
the splitter (pos. 2 on Fig. 2-10) transmits the Z signal that controls the lens scanner placed
in the exchangeable mount. XY socket (pos. 1) transmits X and Y signals that control the
XY-sections of the measuring head scanner.
Fig. 2-10. Z-splitter
1 – XY socket; 2 – Z socket; 3 – CONTROLLER2 socket
The switching cable (Fig. 2-11) supplies the measuring head scanner with the Z signal
from the main controller and the X and the Y signals from the slave controller instead of
the main one.
Fig. 2-11. Switching cable
1 – SCAN+SENSOR socket; 2 – XY socket
17