NT-MDT NTEGRA Spectra Instruction Manual page 27

Inverted configuration with renishaw spectrometer
Hide thumbs Also See for NTEGRA Spectra:
Table of Contents

Advertisement

b. connector of the switching cable (pos. 2 on Fig. 7-2);
ATTENTION! Strictly follow this sequence of connecting the cables!
Otherwise, failure of the instrument is highly probable.
c. connector of the measuring head scanner (pos. 3 on Fig. 7-2).
1 – connector of the exchangeable mount scanner; 2 – connector of the switching cable;
2. Connect the
HEAD
base unit. Connect the stepper motor of the measuring head to the
base unit.
3. Connect the Z-splitter as follows:
CONTROLLER2
connector – to the
XY
connector – to the connector of the lens Z-scanner.
Z
4. Connect the
CONTROLLER1
controller and the
the main controller with proper cables.
5. Connect the spectral unit to the
6. Connect the Andor CCD-camera, the main and the slave controllers to proper interface
cards of the computer.
7. Set the supply voltage switchers at the main controller and the slave controller to the
position corresponding the power supply line (do this only when the controller is
switched off!).
8. Connect the controllers to the power supply.
ATTENTION! Controllers can be connected to an electrical power supply
line of 110/220 V (60÷50 Hz) after setting the voltage selection switch to the
position corresponding to this power supply line. Not following this
instruction may cause damage to the electronic components.
SCAN+SENSOR
Fig. 7-2.
3 – connector of the measuring head scanner
connector of the measuring head to the same-name socket of the
connector – to the same-name socket of the slave controller;
connector of the switching cable;
XY
socket of the base unit to the
socket of the base unit to the same-name socket of
CONTROLLER2
port of the computer.
USB
Chapter 7. Setup and Installation
connectors
SM
socket of the main
HEAD
socket of the
27