Omron KM50 Communications Manual page 83

Smart power monitor
Table of Contents

Advertisement

Section 3 CompoWay/F Communications Data
E2: Alarm History
The history of alarms that occurred in the past can be read.
Up to 10 alarms that occurred in the past can be read out. The smaller the address is, the newer
the information is. If more than 10 alarms occur, the oldest alarms in the history will be deleted.
Data is not saved when power is interrupted. Only the alarms that occur while power supply is ON
can be read out from the history.
The 13 types of alarms that are listed in the following table can be read out. (This includes "No
alarm.")
Table: List of Alarm Types
Read out value
0000 0000 hex
0000 0001 hex
0000 0002 hex
0000 0003 hex
0000 0004 hex
0000 0005 hex
0000 0006 hex
0000 0007 hex
0000 0008 hex
0000 0009 hex
0000 000A hex
0000 000B hex
0000 000C hex
Variable
Address
type
E2
0000
E2
0001
E2
0002
E2
0003
E2
0004
E2
0005
E2
0006
E2
0007
E2
0008
E2
0009
E2
000A
E2
000B
E2
000C
E2
000D
E2
000E
E2
000F
E2
0010
E2
0011
E2
0012
E2
0013
E2
0014
3-42
No alarm
Instantaneous active power upper limit alarm
Instantaneous active power lower limit alarm
Instantaneous regenerative power upper limit alarm
Instantaneous regenerative power lower limit alarm
Current upper limit alarm
Current lower limit alarm
Voltage upper limit alarm
Voltage lower limit alarm
Power factor upper limit alarm
Power factor lower limit alarm
Instantaneous active power lower limit alarm
Instantaneous reactive power lower limit alarm
Parameter
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 1
Date and Time
MMDD hhmm hex*
Cleared
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 2
Date and Time
MMDD hhmm hex*
Cleared
Refer to Table: List of Alarm Types.
Alarm Type
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 3
Date and Time
MMDD hhmm hex*
Cleared
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 4
Date and Time
MMDD hhmm hex*
Cleared
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 5
Date and Time
MMDD hhmm hex*
Cleared
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 6
Date and Time
MMDD hhmm hex*
Cleared
Alarm Type
Refer to Table: List of Alarm Types.
Date and Time
MMDD hhmm hex*
Alarm
of Occurrence
Record 7
Date and Time
MMDD hhmm hex*
Cleared
Alarm type
Set value (monitor value)