804 Series Video Test Generator – User Guide
6

Using the 804 Series Command Line

This section provides procedures on controlling the 804/804A/804B through the command line for testing HDTVs. Refer to the
section entitled,
Command Reference
cable or a USB cable.
6.1

Establishing an RS-232 connection to the 804/804A/804B

The procedures below describe how to connect to the 804/804A/804B through the RS-232 interface for command line control
or control from an automated test system.
1.
Connect an RS-232 cable between your host PC and the 804/804A/804B. The illustration below shows the location of the
RS-232 interface.
2.
(optionally) Connect USB cable between your host PC and the 804/804A/804B on the USB connector on the front. The
illustration below shows the location of the USB interface.
3. Download the .INF file from the Quantum Data website at:
http://www.quantumdata.com/support/804readme.asp#supplemental
http://www.quantumdata.com/support/804Areadme.asp#supplemental
a convenient location on your PC.
4.
Set the USB mode to COM from the Preference menu as shown below. Reboot the 804/804A/804B if you changed this
USB mode from Disk to COM.
Page 20
for a complete list of commands. You can connect from a PC either through an RS-232
or
to your PC and unzip the file. Store it in
Revision B3