Parameterization

7
Parameterization
7.1

Display options

After switching on the auxiliary power, the current process information for the measuring point is
displayed.
In the first line of the display, the current flow direction (
displayed along with the flowrate as a percentage or a physical unit. The second line of the
display shows the totalizer value (7-digit) for the current flow direction, followed by the relevant
unit.
Independent of the pulse factor, the totalizer value is always displaying the current flowrate with
the relevant unit. This indicator is displayed in the following text as process information.
The 4 lines of the display can be customized in the "Display" submenu.
Examples:
Totalizer overflow always occurs at a value of 9.999.999 units. If the totalizer value for a flow
direction is larger than 9.999.999 units, Warning 9 is displayed in the 4th line. The totalizer
software can register up to 250 overflows. The overflow notification can be cleared in the "Reset
totalizer" submenu.
Error condition
When an error occurs, an error or warning message appears in the 4th line of the display.
Flowrate > 103 %
This message is displayed alternatively in plain text and with the relevant error or warning
number. The plain text error message provides the error or warning with the highest priority
only. However, all existing errors and warnings are shown in the number display.
For a list of all possible error messages, see "Error messages".
In addition to the error message, the current output is set to the alarm value (menu "lout for
alarm"). The alarm can be indicated optionally via the contact output, if selected.
Important
Error messages for "enhanced diagnostics" are only indicated on the display or optionally via
the contact output (select "General Alarm" or "ext. Diag. Alarm").
58
V 98.14 l/h
3
V 12.30000 m
3
R 516.0000 m
1. Line
Instantaneous flow forward
2. Line
Totalizer value, forward
3. Line
Totalizer value, reverse (multiplex mode)
FSM4000
F for forward,
R for reverse) is
D184B140U02