ABB FSM4000-SE41F Operating Instruction page 93

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Parameter
Accidental coil ground
DAC max
Magn. linearity
SNR (signal-to-noise
ratio)
D184B140U02
Minimum value
10 MΩ
100
This value depends on the fluid temperature T
T
medium
-40 °C
Factory fingerprint x 0.86
(-40 °F)
-20 °C
Factory fingerprint x 0.90
(-4 °F)
0 °C
Factory fingerprint x 0.93
(32 °F)
20 °C
Factory fingerprint x 0.95
(68 °F)
60 °C
Factory fingerprint x 1
(140 °F)
90 °C
Factory fingerprint x 1.05
(194 °F)
130 °C
Factory fingerprint x 1.09
(266 °F)
180 °C
Factory fingerprint x 1.14
(356 °F)
60 % of the commissioning fingerprint or a
higher value, depends on the application.
Nominal diameter
(1/25 ... 1/12")
DN 3 ... 10
(1/10 ... 3/8")
DN 15 (1/2")
DN 20 ... 25
DN 32 (1 1/4")
DN 40 ... 50
(1 1/2 ... 2")
DN 65 (2 1/2")
DN 100 ... 125
DN 150 (6")
DN 200 ... 300
DN 350 ... 400
DN 450 ... 1000
T
medium
-40 °C
(-40 °F)
-20 °C
(-4 °F)
0 °C
(32 °F)
20 °C
(68 °F)
60 °C
(140 °F)
90 °C
(194 °F)
130 °C
(266 °F)
180 °C
(356 °F)
FSM4000
Parameterization
Maximum value
99.9 MΩ
Design level of the
flowmeter sensor
A
DN 1 ... 2
500
300
400
400
(3/4 ... 1")
400
500
400
DN 80 (3")
500
500
(4 ... 5")
400
400
(8 ... 12")
500
(14 ... 16")
700
(18 ... 40")
.
medium
Factory fingerprint x 0.95
Factory fingerprint x 1
Factory fingerprint x 1.03
Factory fingerprint x 1.05
Factory fingerprint x 1.10
Factory fingerprint x 1.16
Factory fingerprint x 1.21
Factory fingerprint x 1.26
100 %
B
-
300
300
400
500
500
300
300
600
600
700
700
700