ABB FSM4000-SE41F Operating Instruction page 85

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Submenu/parameter
Submenu
Calibration
D184B140U02
Value range
As Startup
Accept
Yes, exit via C/CE
Yes → Enter
Submenu
Coil Temp
Temp. in °C
-100 ... 100
0.0
Submenu
CabLength
Length [m]
0 ... 200
0.0
Submenu
Temp. Offset
Temp. in °C
-100 ... 100
0.0
Import RT20
Accept
Yes, exit C/CE
New 500
ohms
Old 450 ohms
Yes → Enter
Electr. Zero Point
Electr. Zero
Yes, exit C/CE
Point
New 150 180 µV
Old 200 190
µV
Yes → Enter
Electr. Analysis
N1: 12 µV 286°
Yes, exit C/CE
N2: 10 µV 283°
A1: 650 µV 133°
A2: 670 µV 135°
Comment
Copies the current data record in the Fingerprint "as
Startup" (commissioning fingerprint).
Enter the current coil temperature to readjust the coil
resistance at 20 °C. Perform adjustment only on a "cold"
device. See also chapter "7.4 Additional information
regarding use of enhanced diagnostic functions".
Enter the length of the signal cable.
Note: Information must be entered to calculate the coil
resistance / coil temperature.
Offset correction for temperature display.
Import the reference resistance for the coil.
See also chapter "7.4 Additional information regarding
use of enhanced diagnostic functions".
Note: This menu is displayed only when the coil
resistance or coil temperature are enabled as an error or
warning message.
This adjustment is required in order to use the diagnostic
measurement values for the electrode voltage E1/E2 and
the electrode balance as well as the wiring errors for the
electrode state.
Display the current (new) and stored (old) zero values for
electrode E1/E2.
Select "Enter" to import the current zero points
Note: The fluid in the flowmeter sensor must be at
absolute standstill.
Displays stored electrode zero and current electrode AC
voltage acc. to amount and phase.
Where
N1: Current measured electrode AC voltage E1
N2: Current measured electrode AC voltage E2
A1: Stored zero E1
A2: Stored zero E2
FSM4000