Connection Examples For Peripherals (Incl. Hart) - ABB FSM4000-SE41F Operating Instruction

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Electrical connections
5.5

Connection examples for peripherals (incl. HART)

Current output
I
E
+31
0/4 ... 20 mA
0/2 ... 10 mA
-32
Current output:
selectable in software
Active
Function:
0/4 ... 20 mA (0 Ω ≤ RB ≤ 560 Ω)
0/2 ... 10 mA (0 Ω ≤ RB ≤ 1120 Ω)
Terminals: 31, 32
Fig. 40:
Current output active with/without HART protocol (4 ... 20 mA)
Contact output
I
E
R
+U
41
42
U
*
CE
R
B
I
CE
Contact output:
selectable in software
Passive
Function:
"closed" 0 V ≤ U
"open" 16 V ≤ U
Terminals: 41, 42
Fig. 41:
Contact output for system monitoring, Max.-Min. alarm for empty tube or forward/reverse signal
Contact input
I
E
81
R
i
82
Contact input:
selectable in software
Passive
Function:
"On" 16 V ≤ U
"Off" 0 V ≤ U
Ri = 2 kΩ
Terminals: 81, 82
Fig. 42:
Contact input for external totalizer reset and external zero return
44
G00243
(with HART only 4 ... 20 mA)
*
B
0 V
24 V DC
G00244
≤ 2 V, 2 mA ≤ I
CEL
CEL
≤ 30 V, 0 mA ≤ I
CEH
CEH
+ 16 ... 30 V DC
0 V
G00245
≤ 30 V
KL
≤ 2 V
KL
I = internal
E = external
≤ 220 mA
≤ 2 mA
FSM4000
Note
Horn and alarm light are shown only
as examples. Other suitable devices
such as bells, sirens, buzzers, etc.,
can also be used.
D184B140U02

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsm4000-se21Fsm4000-se21fFsm4000-s4Fsm4000