ABB FSM4000-SE41F Operating Instruction page 74

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Parameterization
Submenu/parameter
Submenu
Pulse output
Qmax Pulses
3
2.5 m
/h
Pulse Factor
100/m
3
Pulse width
0.1 ms
Pulse (Act./Pas.)
Active
Submenu
Switched contacts
Contact input
No function
Contact output
No function
74
Value range
Comment
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Numerical entry
Expanded flow totalization to the maximum flow range
Qmax DN. This parameter is displayed only if the
measuring range specified is less than Qmax DN.
Numerical entry
The pulse width is configurable only for the limits
calculated by the transmitter. The limits are based on the
Dynamic limits
pulse factor. The transmitter cannot leave the max. limits
of 2000 ms and 0.1 ms.
Numerical entry
The pulse width is configurable only for the limits
calculated by the transmitter. The limits are based on the
Dynamic limits
pulse factor. The transmitter cannot leave the max. limits
of 2000 ms and 0.1 ms.
Display only
Displays the "active" or "passive" jumper position (see
page 47).
No function
The contact input function is not displayed for PROFIBUS
Totalizer reset
PA or FOUNDATION Fieldbus.
Select the function for the contact input terminals 81/82.
Ext. cut-off
"Totalizer Reset" resets the internal totalizer and overflow
Ext. Syst. NP.
values when a switching voltage is connected. For "Ext.
Zero Return" all outputs are switched off, if a switching
voltage is connected. Example: Undefined filled pipeline
condition after turning off pumps or cyclic cleaning or
flushing procedures whose flowrates must be hidden.
Select "Ext. Syst. NP" to adjust the system zero point.
The fluid must be at absolute standstill during the
adjustment. The message "Warning 13" is displayed
during the adjustment.
No function
The switching behavior for the contact outputs (terminal
41/42) can be selected as normally open (/_ _) or
F/R signal /_
normally closed (_ _). For the "General Alarm" all
F/R Signal __
detected errors (0 - 6, A, B, C, D, E, F, G) are signaled via
General Alarm /_
the terminals. Only diagnostic errors are signaled when
General Alarm _ _
"ext. Diag. Alarm" is selected. If you want to report a low
Max/Min Alarm/_
flow condition, select "Min Alarm". Similarly, a high flow
Max/Min Alarm _ _
condition is signaled with "Max Alarm". If the signal is set
Min Alarm /_
for "Max/Min Alarm", then a signal is sent over the
Min Alarm _ _
terminals when the flowrate is above or below the range
Max Alarm /_
between the "Max-Alarm" value and the "Min-Alarm"
Max Alarm _ _
value, i.e., when the flowrate is larger than the "Max-
Empty Pipe /_
Alarm" value or less than the "Min-Alarm" value.
Empty Pipe _ _
Ext. Diag. Alarm /_
Ext. Diag. Alarm __
FSM4000
D184B140U02