Spare Parts List; Fuses For Transmitter Electronics - ABB FSM4000-SE41F Operating Instruction

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Spare parts list

10 Spare parts list
Important
Please contact Customer Center Service acc. to page 2 for nearest service location.

10.1 Fuses for transmitter electronics

Warning – Electrical voltage risk!
When the housing is open, EMC protection is impaired and protection against contact is
suspended.
Power to all connecting cables must be switched off.
Replacing the fuse
1. Switch off supply power. After waiting 40 minutes, remove the housing cover from the field
housing unit.
2. To replace the fuses for the remove the protective cover (1), loosen and remove the 3
screws (2).
Fig. 54
1 Protection cover
2 Screws for cover
3
4
5
108
1
Parameter
Pre fuse for supply power on backplane (4 A-T)
Fuse for supply power on the transmitter electronic unit
Fuse for 24 V AC/DC (4 A-T)
Fuse for 100 -230 V AC (1 A-T)
Driver circuit fuse F103 (0.8 A)
2
G00579
3 Pre fuse for supply power on the
backplane
4 Fuse for supply power on the transmitter
electronic unit
5 Fuse for driver circuit on the backplane
FSM4000
5
4
Order number
D151B003U08
D151B003U08
D151B003U05
D151F003U18
D184B140U02
3
G00581

Advertisement