Flowmeter Sensor In Stainless Steel Design Model Se21; Ground For Devices With Protective Plates; Ground With Conductive Ptfe Grounding Plate - ABB FSM4000-SE41F Operating Instruction

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Mounting
4.8.5

Flowmeter sensor in stainless steel design model SE21

Ground the stainless steel model as shown in the figure. The measuring fluid is grounded via
the adapter (1) and an additional ground is not required.
Fig. 29
4.8.6

Ground for devices with protective plates

The protective plates are used to protect the edges of the liner in the measuring tube, e.g., for
abrasive fluids. In addition, they function as a grounding plate.
For plastic or pipes with insulating lining, electrically connect the protective plate in the same
manner as a grounding plate.
4.8.7

Ground with conductive PTFE grounding plate

For devices with a meter size between DN 10 ... 250, grounding plates made of conductive
PTFE are available. These are installed in a similar way to conventional grounding plates.
32
1
FSM4000
G00533
D184B140U02

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsm4000-se21Fsm4000-se21fFsm4000-s4Fsm4000

Table of Contents