Displaying The Diagnostic Values - ABB FSM4000-SE41F Operating Instruction

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Parameterization
7.4.3

Displaying the diagnostic values

7.4.3.1
Measurement values
A maximum of four diagnostic values can be displayed (Fprt1, Fprt2, Fprt3, Fprt4). Values can
be selected in the submenu Diagnostics / Fingerprint / Fprt1 - 4. In addition to the type of
measurement value, the desired data record has to be configured. Through the combination of
these two settings, every stored measured value can be displayed. To differentiate between the
values in the display an identification code (ID) is prefixed before to each value.
As a result, each line appears as follows:
e.g., CV_RCL 80.5 ohm for the current coil resistance.
Data record
CV_: Current value
H1_: History 1
H2_: History 2
H3_: History 3
H4_: History 4
H5_: History 5
H6_: History 6
H7_: History 7
H8_: History 8
H9_: History 9
FV_: Factory Fingerprint
DF_: Startup Fingerprint
P1_: Product 1-Fingerprint
P2_: Product 2-Fingerprint
94
CV _ RCL 80.5 ohm
Type
UAC: Coil voltage (AC)
IAC: Coil current (AC)
RCL: Coil resistance
TCL: Coil temperature
RIS: Insulation resistance
UE1: Electrode voltage E1
UE2: Electrode voltage E2
BAL: Electrode balance
LIN: Magnetic linearity
SNR: Signal-to-noise ratio
DAC: Max. DAC value
FSM4000
Numerical value
Unit
X.X
V
X
mA
Value > 100 X
Ohm
Value < 100 X.X
X
°C
Value > 100 X
MOhm
Value < 100 X.X
X
µV
X
µV
E1 ≥ E2 → +X
%
E1 < E2 → -X
%
X
%
X
%
X
Dig =
Digits
D184B140U02