ABB FSM4000-SE41F Operating Instruction page 96

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Parameterization
The error registers for diagnostic values are structured as follows:
ID code
_______________________________
ID
code
HMx
HMi
EMx
EMi
WMx
WMi
The information
voltage and electrode balance.
The error registers for wiring errors are structured as follows:
ID code
_______________________________
ID code
ConWarn
ConErr
96
Register name
Hist Max Errors
Hist Min Errors
Curr Max Errors
Curr Min Errors
Curr Max Warn
Curr Min Warn
EMx _R______B__ indicates Max. Alarm in the current values for the coil
Register name
Wiring warning
Wiring Error
Content of register
Error / Warning: (Max.) values overshot while transmitter is in
operation.
Error / Warning: (Min.) values undershot while transmitter is in
operation.
Error: (Max.) present available current overshoots.
Error: (Min.) present available current undershoots.
Warning: (Max.) present available current overshoots.
Warning: (Min.) present available current undershoots.
Content of register
Current warnings for wiring
Current errors for wiring
FSM4000
Set
Off
Description
S
Signal quality
L
Magnetic linearity
B
Electrode balance
e
Electrode voltage E1
E
Electrode voltage E2
D
DAC value
I
Insulation resistance
T
Coil temperature
L
Coil resistance
R
Coil voltage
C
Coil current
Set
Off
Description
M
Coil short-circuit to
ground
E
Wiring error for
electrodes
D184B140U02

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fsm4000-se21Fsm4000-se21fFsm4000-s4Fsm4000

Table of Contents