Operating S4 with an older sensor

11 Operating S4 with an older sensor
When operating a transmitter with an older model, the FRAM must be used with the values Cs =
100% and Cz = 0%. If the transmitter has been set up in this way and the devices have been
mounted according to the connection diagram (see page
appears after switching on the supply power. Press any key to enter the following information. If
you need to correct your entry, you can do so afterward under the system data submenu
"Flowmeter Primary" in "Primary Setup". See page 67.
Submenu
Primary Setup
Important
If no values are listed for the reference voltage on the name plate, you can request this
information from ABB Service. Please provide order number with your request.
e-mail:
Phone: +49 180 5222 580
112
Key
Main frequency
50 Hz
Type of primary
10DS3111 E
Meter size
100 mm 4 in
Reference voltage
70 mV
Meter factor
1.0
parts-repair-goettingen@de.abb.com
Entries can be made one after another. In
the submenu "Primary Setup", select the line
frequency at which the flowmeter sensor
was calibrated. See name plate. Select the
flowmeter sensor type from the list in the
table. Check that the letter after the model
number and the nominal diameter range are
correct. Otherwise, the transmitter will not
access the correct reference data.
Enter the relevant meter size. The values
?
DN 300
are listed for DIN and ANSI sizes. Enter the
reference voltage (see the name plate on
the flowmeter sensor) in numerical format.
(For flowmeter sensor 10D1422 or
10DI1425 > DN500, the reference voltage is
not required.)
Enter the meter factor (for model
10D1462/72 only). See name plate.
Select the measuring range and other
process parameters (damping, low cut-off,
current/pulse output, display format, etc.). A
system zero adjustment must be performed
G00585
with a full pipeline at standstill (0 m/s).
FSM4000
40
and on), the "Primary Setup" menu
D184B140U02