Parameterization
Submenu/parameter
80
Value range
Submenu
Alarm Coil Current
Min. value
0 ... 500
20.0 mA
Max. value
0 ... 500
500.0 mA
Message
Off, warning, error
Off
Submenu
Alarm Coil Resist
Min. value
0 ... 1500
2 ohms
Max. value
0 ... 500
500 ohms
Message
Off, warning, error
Off
Submenu
Alarm Coil Temp
Min. value
-100 ... 200
-50 °C
Max. value
-100 ... 200
200 °C
Message
Off, warning, error
Off
Comment
Set the time-out limits and alarm behavior for values over
or under the coil AC.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Set the time-out limits and alarm behavior for values over
or under the coil resistance.
Note: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see
"Submenu Adjustment" on page 85.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Note: If coil resistance is being measured (at least every
60 s), the most recently acquired flowrate is retained
during the measurement (2 s approx.).
Set the time-out limits and alarm behavior for values over
or under the coil temperature.
Note: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see
"Submenu Adjustment" on page 85.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Note: If coil resistance is being measured (at least every
60 s), the most recently acquired flowrate is retained
during the measurement (2 s approx.).