Summary of Contents for ABB CoriolisMaster FCM2000
Page 1
Operating Instruction Mass Flowmeter D184B111U02 CoriolisMaster FCM2000 Pos: 1 /Titelblätter / Copyright/BA-IA/Durchfluss/FCM2000 @ 6\mod_1171653775140_3101.doc @ 67895 @ Standard Software D699G001U01 B.3x D699G001U02 A.1x D699G001U03 A.1x ® P R O F I PROCESS FIELD BUS B U S Pos: 2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @...
Read and follow the instructions in this manual. • Observe the technical ratings; refer to the section “Technical limit values”. Use only allowed liquids for measurement; refer to the section “Allowed fluids”. • Pos: 5.6 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
Pos: 5.8 /Sicherheit/Allgemein/Hinweis zur bestimmungswidrigren Verwendung (Wartung/Reparatur) @ 0\mod_1129707002440_3101.doc @ 3240 @ Repairs, alterations, and enhancements, or the installation of replacement parts, are only permissible insofar as these are described in the manual. Approval by ABB Automation Products GmbH must be sought for any activities beyond this scope. Repairs performed by ABB-authorized specialist shops are excluded from this.
Using the device in a manner that does not fall within the scope of its intended use, disregarding this manual, using underqualified personnel, or making unauthorized alterations releases the manufacturer from liability for any resulting damage. This renders the manufacturer's warranty null and void. Pos: 5.13 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
This symbol indicates operator tips, particularly useful information, or important information about the product or its further uses. It does not indicate a dangerous or damaging situation. Pos: 5.17 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
40.375 Cal.: V: +/-0.4 % D: +/- 5 g/l Qmax DN: 100 kg/min Tmed: -50 °C…+180 °C ABB Automation Products GmbH 37070 Göttingen – Germany G00352 Fig. 1 1 Order number 7 Meter pipe material 2 Complete model number 8 Calibration factor...
Page 11
1.4571 72.753 Cal.: F: +/-0.15% D: +/- 5 g/l Qmax DN: 920 kg/min TAG no.: FT–402 ABB Automation Products GmbH 37070 Göttingen – Germany G00353 Fig. 2 1 ATEX approval 6 Meter size and protection class 2 ATEX temperature classes...
Page 12
S.-Nr.: 0012345 0045 DN 50 / PN 40 Material: 1.4571 Manufactured: 2002 PED: Fluid 1, Gas ABB Automation Products GmbH 37070 Göttingen - Germany G00410 Fig. 4 The factory plate contains the following information: 1 CE mark (with number of labeled location) to confirm the device meets the requirements of pressure equipment directive 97/23/EC.
Prior to using corrosive and abrasive materials for measurement purposes, the operator must check the level of resistance of all parts coming into contact with the materials to be measured. ABB Automation Products GmbH will gladly support you in selecting the materials, but cannot accept any liability in doing so.
If it is not possible to dispose of old equipment properly, ABB Service can accept and dispose of returns for a fee. Pos: 5.29 /Sicherheit/Durchfluss/FCM2000/Sicherheitshinweise zum Transport (Durchfluss) @ 7\mod_1174894735718_3101.doc @ 74253 @ 2 1.12 Safety instructions for transport...
The purpose of these screws is to allow for proper disposal of any contaminated fluids (in the event of pipe leakage). Inspections screws may not be used under any circumstances to connect trace heating. D184B111U02 CoriolisMaster FCM2000...
Pos: 8 /Aufbau und Funktion/Durchfluss/FCM2000/Aufbau @ 6\mod_1171654750218_3101.doc @ 68021 @ 2 Design The ABB Automation Products Mass Flowmeter operation is based on the Coriolis principle. The MC2 construction uses the classical parallel meter pipes and is characterized, in particular, by a space saving and rugged design, a wide flowmeter size spectrum at an advantageous price to the customer.
Page 18
Simplified Representation of the Coriolis Forces ω × = -2 m ( = Coriolis force ω = Angular velocity ν = Velocity of the mass m = Mass Pos: 10 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
Self-monitoring, diagnostics On-site display/totalization Field optimized flow/density Protection class acc. to EN ME2: IP 65 / 67, NEMA 4X 60529 MC _ _ : IP 67, NEMA 4X Pos: 13 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
The mass meter does not require any flow conditioning inlet straight G00361 sections. Care should be exercised to ensure that any valves, gates, Fig. 12: Vibrations sight glasses, etc., do not cavitate and are not set into vibration by the flowmeter sensor. D184B111U02 CoriolisMaster FCM2000...
Tighten the cover fittings. • Install the separate transmitter at a largely vibration-free location. • Do not expose the transmitter to direct sunlight. Provide appropriate sun protection as necessary. Pos: 28.3 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
3. Ensure that the pressure relief valve is correctly installed and properly tightened, so the gasket is seated correctly. Whenever it is removed the gaskets should be replaced. G00370 Fig. 25 Connecting nipple Connecting nipple Pos: 28.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Loosen the mounting screw (approx. 2 turns) Rotate converter to desired orientation Tighten the mounting screws G00381 Fig. 26: Rotating the Converter Housing Important After positioning the converter it is essential that a head set screws are tightened. CoriolisMaster FCM2000 D184B111U02...
4. Check that the gasket is seated properly. 5. Carefully reinstall the housing cover in the new position. Only then will Protection Class IP 67 (NEMA 4X) be maintained. Pos: 28.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
4 Cover for the supply power. For supply power connections, see Fig. 41 Important Please lay the cable shield of the signal cable under the collecting bracket in the connection area! Pos: 28.15 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Tighten the four screws with a 4 mm hexagon socket wrench to secure the adapter. Install the multi-plug connector and tighten the screws on the plug to ensure proper seal. G00412 Pos: 29 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
• Inductance: ca. 1 mH/km • Max. cable length: 50 m (164 ft.) Cut to length and terminate the cable as shown. Important Use wire end sleeves. 80 (0,59) 65 (2,56) (0,59) 8 (0,31) G00396 Fig. 30 D184B111U02 CoriolisMaster FCM2000...
2. Cut the woven shield to a length of approx. 15 mm (0.59”). 3. Remove cable drain wire and foil shield. 4. Remove cable insulation and attach cable conductor sleeves. 5. Wind shield core around woven shield. CoriolisMaster FCM2000 D184B111U02...
Page 37
G00413 Fig. 33 1 Make sure during installation that the cable is provided with a water trap (1). For vertical installation, align the cable glands pointing downward. Pos: 31.2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Fig. 36: Contact input for external totalizer reset and external zero return Pulse Output active passive, optocoupler + 16 ... 30 V DC 24 V+ G00246 G00247 Fig. 37: Pulse output active and pulse output passive, optocoupler CoriolisMaster FCM2000 D184B111U02...
Page 43
I = 14 mA (normal operation) I = 26 mA (in the event of error/FDE) Terminals: 97 / 98 Connection example via M12 plug, see Fig. 39 Pos: 31.4 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
If the ground wire (PE) in connection box of flowmeter is connected it has to be assured that no dangerous potential difference between the ground PE and potential balance can show up in the explosion hazardous area. Pos: 31.7.4 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
For additional information, see the separate data link description. 1.2 (DSV401 R2), respectively. Other desired tool/system inte- grations (e.g., AMS or Siemens S7) are available upon request. DSV401 communication tool for HART, free 90-day test version also available upon request. DTMs are included in DSV401. D184B111U02 CoriolisMaster FCM2000...
ABB at no cost. connection The files needed for operation can also be downloaded at www.abb.com/flow -> Coriolis Massflow -> Fieldbus Files Bus topology Pos: 36 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829 @ •...
If the data is not identical, the transmitter data is replaced automatically. Once completed, the message "Ext.Dat.loaded" is displayed. The measuring equipment is now ready for operation. The display shows the current flowrate. Pos: 38.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
“automatic” and select ENTER to start the adjustment. You can choose between slow or fast adjustment. Slow adjustment generally provides a more accurate zero point. Important All parameters are stored automatically in the FRAM Pos: 38.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
Switch 9 and A have no significance for the address setting. Example: Address 50 set by switch: 50 dec = 32 hex = 110010 binary → Switch 2, 5, 6 and 8 1 2 3 4 5 6 7 8 9 A G00363 Fig. 50 D184B111U02 CoriolisMaster FCM2000...
“Error – operating protection” and the input will be rejected. It is also possible to use a cover locking screw with a hole to seal the compact unit so that parameter changes cannot remain undetected. Pos: 38.15 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
If the signal outputs are to be connected to intrinsically safe circuits, it is recommended that the included light blue caps be used for the corresponding cable connectors. Pos: 38.21 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
(Fig. 54). See also chapter Interconnection Diagrams. Plug-in jumper position Standard configuration NAMUR configuration preferred for Ex „e“ preferred for Ex „i“ (Default configuration) G00368 Fig. 54: Plug-in jumper positions Pos: 38.24 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Page 56
“intrinsically safe” or “non-intrinsically safe”. A combination is not permitted. The test voltage for the output circuits for both Types of Protecton is U = 60 V. Pos: 38.27 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
The pipeline and the flowmeter sensor insulation should be installed as shown. The max. insulation thickness at the flowmeter sensor is 100 mm (4“). G00331 Fig. 57 max. 100 mm (4“) Pos: 38.30 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
• The connections between the flowmeter sensor and transmitter may only be made with the cable supplied by ABB Automation Products. • For safe operation it is essential to follow the instructions in this Operating instruction. •...
Access the parameter to be changed and set the new, selected or default parameter. The ENTER function is only active for 10 seconds. If no entries are made during this period, the old value is displayed on the converter. CoriolisMaster FCM2000 D184B111U02...
Page 61
There are two different methods of entering data: • Numerical input • Entry from specified table Important When entering data, the values are checked for plausibility and, if necessary, rejected with an appropriate message. Pos: 41.2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
STEP or DATA "Progr. level" or meter size Technician ENTER "Progr. level" blocked End point for "process information" C/CE → V 98.14 % (converter remains online). →V 12,000 m Pos: 41.4 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
"Technician":"Technician":Expanded standard menus with the complete set of user specific parameters "Service":Additional set of Service menus only available to ABB Automation Products Service personnel. If the Prog. Prot.-Code is set to 0 (default setting) the programming levels "Standard" or "Technician" can be accessed without entering a code number.
Page 64
Note DensiMass code If ordered with DensiMass option, this menu shows the order specific DensiMass code. If the 4289794 DensiMass shall be activated after shipping the meter, please contact your ABB service or sales organisation. DensiMass Code invalid Submenu Concentration...
Page 65
Calculated values are marked with a "C". Rules for valid values see 6.4.3. Matrix calculation Calculation of matrix based on above entered values. Missing values are inter- or extrapolated. Enter matrix Store or discard of above values. finish D184B111U02 CoriolisMaster FCM2000...
Page 66
Unit name 3 ASCII ASCII In this menu the name or abbreviation of the programmable mass flow unit can be entered. The Technician* name can have a maximum of 3 characters. Not contained in the fieldbus software versions. CoriolisMaster FCM2000 D184B111U02...
Page 67
This menu displays the maximum mass flowrate for the selected instrument size. 100.00 kg/min Order no. 16 ASCII Display Display of the Order Number. It is also listed on the Name Plate and on the label on the external 240012345X004 characters memory module. D184B111U02 CoriolisMaster FCM2000...
Page 68
The concentration accuracy can be optimized in the field to come close to the repproducibilty of the 0.00 % Technician meter. The correction value has to be entered in the selected unit. C correction tab. 2 0.00 % CoriolisMaster FCM2000 D184B111U02...
Page 69
The Max-Alarm value must be larger that the associated Min-Alarm value. Min Alarm Qm 0 ... 105 % float Set the lower mass flowrate limit. It must be less than the upper mass flowrate limit. 0.00 % Technician* Minimum: Maximum: 105 % D184B111U02 CoriolisMaster FCM2000...
Page 70
Lower concentration limit must be lower than the Max alarm limit. 0.00 % Technician Max Alarm Concentr. -5 % ... 105.0 % The max alarm concentration limit must be bigger than the Min alarm limit. 0.00 % Technician CoriolisMaster FCM2000 D184B111U02...
Page 71
TAG Number Totalizer Mass Totalizer Mass → F Totalizer Mass ← R Totalizer Volumes Totalizer Vol. → F Totalizer Vol. ← R Totalizer Net Mass Total. Net Mass → F Total. Net Mass ← R Pipe frequency Blank D184B111U02 CoriolisMaster FCM2000...
Page 72
TAG Number Totalizer Mass Totalizer Mass → F Totalizer Mass ← R Totalizer Volumes Totalizer Vol. → F Totalizer Vol. ← R Totalizer Net Mass Total. Net Mass → F Total. Net Mass ← R Pipe frequency Blank CoriolisMaster FCM2000 D184B111U02...
Page 73
Display of totalizer status for reverse direction; displayed in the Forward/Reverse operating mode 123456.78 l Input Overflow ← R Display Number of totalizer overflows Max. of 65535 overflows 1 overflow = 10000000; displayed in the Forward/Reverse operating mode D184B111U02 CoriolisMaster FCM2000...
Page 74
Reset totalizer values, all flow totalizers are simultaneously reset to zero. The sequential Reset ? display of each totalizer is shown to certify that all Yes → ENTER totalizers have been reset. CoriolisMaster FCM2000 D184B111U02...
Page 75
Submenu Not contained in the fieldbus software versions. Current output 1 Output float The selections in this menu are used to defined for which variable current output 1 is to be based. Density Temperature Concentration Qm Concentration D184B111U02 CoriolisMaster FCM2000...
Page 76
100 % must be at least 10 °C larger than the temperature value for a current output value of 0 %. Qm% → l = 100 % Only displayed for "Output of Qn Conc." Net massflow at current output 100 %. 120 kg/min CoriolisMaster FCM2000 D184B111U02...
Page 77
The setting for the alarm current 21 mA Technician* is independent of the current output range selected since all range end values are 20 mA. The High-Alarm-Current can be set between the limits of 21 mA and 26 mA. D184B111U02 CoriolisMaster FCM2000...
Page 78
Minimum: -50 °C Maximum: 180 °C The temperature value for a current output value of 100 % must be at least 10 °C larger than the temperature value for a current output value of 0 %. CoriolisMaster FCM2000 D184B111U02...
Page 79
2 can be selected. • Ext. zero return (Current and pulse outputs are set to 0% flowrate. Internal totalizers are halted.) • Totalizer reset. (resets all mass and volume flow totalizers) Not contained in the fieldbus software versions. D184B111U02 CoriolisMaster FCM2000...
Page 80
(Packed ASCII). Descriptor 16 ASCII HART descriptor 123456789ABCDE The length is limited to 16 numbers and/or capital letters (Packed ASCII). Date 1.1.1900 ... HART file 07.07.2001 31.12.2155 Unit number 0 ... 999999 Display 1234567 CoriolisMaster FCM2000 D184B111U02...
Page 81
TB TotVol > V • Temperature TB TotVol < R • Internal mass totalizer (in contrast to the PA TB Volume Flow totalizer) • Internal volume totalizer (in contrast to the PA totalizer) Only contained in the PROFIBUS PA software. D184B111U02 CoriolisMaster FCM2000...
Page 82
TB Mass Flow can be assigned to the TOT 1 channel, and TB Volume Flow to the TOT 2 TB Volume Flow channel. These PROFIBUS PA totalizers can deviate from the FCM2000 internal totalizers. Only contained in the PROFIBUS PA software. CoriolisMaster FCM2000 D184B111U02...
Page 83
A current between 0 ... 26 mA can be simulated. (In version A.00 the entry is made in percent) Iout 1 Not contained in the fieldbus software. 10 mA D184B111U02 CoriolisMaster FCM2000...
Page 84
Simulated values for the variables can be entered which exceed the alarm limits to test the operation of the alarms. The settings in the simulation menu are not stored. After a power interruption all simulation menus are turned off. CoriolisMaster FCM2000 D184B111U02...
Page 85
Temp. housing Measure float 20 °C Enter Enter Technician* Function test Technician* The two HART-Frequencies (1200 Hz and 2200 Hz) can be selected and transmitted. HART Transmitter Function test Technician* The received HART commands are displayed. HART command D184B111U02 CoriolisMaster FCM2000...
Page 86
Example Display Ext. Cut-off Power outage Display the number of power outages since the last reset. Status The error and warning registers as well as the mains interrupt counter can be reset. Reset Reset Yes → ENTER CoriolisMaster FCM2000 D184B111U02...
Page 87
(B.30) are displayed. In addition to the software identification in the operator menu, the identification can also be found on the information tag on the converter module. Pos: 41.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Display of Totalizer Net Massflow according to net-massflow Totalizer Net Mass → F Display of Totalizer Net Massflow - Forward Totalizer Net Mass ← R Display of Totalizer Net Massflow - Reverse Display the frequency of the meter pipe Pipe frequency Blank line CoriolisMaster FCM2000 D184B111U02...
Page 89
Display of the respective state of the Transducer Block FB TOT2 Out Display of the respective PA totalizer value within Function Block FB TOT2 Status Display of the respective state of the Transducer Block only in Technician menu D184B111U02 CoriolisMaster FCM2000...
0.2 ms (5 kHz) → maximum pulse width = Period /2 = 0.1 ms → Result: The entered pulse factor and pulse width were both automatically decreased Pos: 41.10 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
Concentr. 1 <…< Concentr. x < … < Concentr. N for 1 ≤ x ≤ N Concentr. 1 >…> Concentr. x > … > Concentr. N for 1 ≤ x ≤ N D184B111U02 CoriolisMaster FCM2000...
Page 92
60 °BRIX 1.101 kg / l 1.018 kg / l 1.011 kg / l The unit of density and temperature corresponds to the chosen units in submenu units. Pos: 41.13 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
Parameterization Pos: 41.14 /Konfiguration, Parametrierung/Durchfluss/FCM2000/Software-Historie @ 7\mod_1171894144875_3101.doc @ 69131 @ Software history According to NAMUR recommendation NE53, ABB offers a transparent and traceable software history. 6.5.1 Standard and HART version Software D699G001U01 Software Revision date Type of changes Documentation Version A.1x...
Important Please note that the new 8 kB FRAM external storage medium can only be read as of software version A40 (or A10 of the fieldbus software)! Pos: 42 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
0% and the current output, which signals the density, would be set to Low-Alarm regardless of the settings in the current output menu. Pos: 44.2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Flowrate is in the reverse direction Operating mode set to forward, In Submenu „Operating mode” actual flowrate is in reverse change flow direction to Reverse Q Forward/Reverse Pos: 44.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Pt 100 is defective Check the resistance of PT100 on of the measurement variable Qm the primary flowmeter a density of 20°C is used, i.e. for a fluid temperature near 20 °C the measurements will be correct CoriolisMaster FCM2000 D184B111U02...
Page 99
Concentration given and chosen matrix Error Parameters cannot be changed The operating protection switch is Deactivate the hardware active protection switch. prot. user trans Pos: 45 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
For repairs to the lining, electrodes or magnet coil, the flowmeter must be returned to the local office of ABB. Important When sending the flowmeter sensor to the local office of ABB, complete the return form in the appendix and include with device. Pos: 46.5 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @...
Warning – Electrical voltage risk! When the housing is open, EMC protection is impaired and protection against contact is suspended. • Power to all connecting cables must be switched off. Pos: 47 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Cover small (rear view) Fig. 60: Replaceable parts list transmitter field-mount housing Name Order number Transmitter module Please contact your ABB service team. Connection board standard D685A1020U10 Cover large, complete D641A030U01 Pan head Phillips screw M 3x5 DIN 7985 stn stl...
Page 103
Spare parts list G00385 Fig. 61: Fuse in field-mount housing 1 Fuse field-mount housing G00386 Fig. 62: Fuse in transmitter module 1 Fuse in module Pos: 50 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
± 0.4% of measured value ± 0.05 % of Q ± 0.25% of measured value ± 0.05 % of Q ± 0.15% of measured value ± 0.05 % of Q ± 0.1% of measured value ± 0.05 % of Q CoriolisMaster FCM2000 D184B111U02...
(EN 61326) as well as low voltage directive 73/23/EC (EN 61010-1). Pos: 54 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829 @ Change from one to two columns Pos: 55 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ CoriolisMaster FCM2000 D184B111U02...
FM Approvals (US) Pos: 56.6 /Hinweise/Wichtig/Wichtig - Verweis auf Download-Bereich @ 21\mod_1215086830219_3101.doc @ 204693 @ Important All documentation, declarations of conformity, and certificates are available in ABB's download area. www.abb.com/flow Pos: 56.7 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @...
Page 125
Which substances have come into contact with the device? We hereby state that the devices/components shipped have been cleaned and are free from any dangerous or poisonous substances. Town/city, date Signature and company stamp Pos: 57 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @ D184B111U02 CoriolisMaster FCM2000...
Pressure loss curves..........107 Exchanging the transmitter ......49, 101 Pressure relief valve ..........29 Explosion protection data ........44 Repairs, alterations, and enhancements ....7 Factory plate ............12 Representation of the Coriolis Forces .....18 Factory Tag..............10 Returning devices ............13 Flowmeter sensor ..........100 CoriolisMaster FCM2000 D184B111U02...
Page 127
WEEE Directive ............14 Target groups and qualifications......13 Pos: 61 /==== Wechsel zwei- auf einspaltig ==== @ 0\mod_1130421955859_3101.doc @ 3829 @ Change from one to two columns Pos: 62 /Rückseiten/Göttingen @ 0\mod_1138784575421_3101.doc @ 3125 @ ===== Ende der Stückliste ===== D184B111U02 CoriolisMaster FCM2000...
Need help?
Do you have a question about the CoriolisMaster FCM2000 and is the answer not in the manual?
Questions and answers