ABB FSM4000-SE41F Operating Instruction page 81

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000-SE41F:
Table of Contents

Advertisement

Submenu/parameter
D184B140U02
Value range
Submenu
Alarm Ins Resist
Min. value
0 ... 100000
1 megaohm
Max. value
0 ... 100000
50 megaohm
Message
Off, warning, error
Off
Submenu
Alarm E1 Voltage
Min. value
0 ... 3000
0.0 µV
Max. value
0 ... 3000
3000.0 µV
Message
Off, warning, error
Off
Submenu
Alarm E2 Voltage
Min. value
0 ... 3000
0.0 µV
Max. value
0 ... 3000
3000.0 µV
Message
Off, warning, error
Off
Comment
Set the time-out limits and alarm behavior for values over
or under the coil temperature.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Note: If coil insulation resistance is being measured (at
least every 10 minutes), the most recently acquired
flowrate is retained during the measurement (2 s approx.).
Set the time-out limits and alarm behavior for values over
or under the electrode AC voltage for electrode E1.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Note: To properly display the electrode voltage, the
electrode zero must be adjusted. For additional
information, see "Submenu Adjustment" on page 85.
Set the time-out limits and alarm behavior for values over
or under the electrode AC voltage for electrode E2.
Lower range limit for alarm.
Upper range limit for alarm.
Defines the alarm behavior.
Note: To properly display the electrode voltage, the
electrode zero must be adjusted. For additional
information, see "Submenu Adjustment" on page 85.
FSM4000
Parameterization
81